Eco-friendly and high-performance, the U-5100 is the best choice for a wide range of applications from teaching and routine QC to biotech and environmental testing.
This entry-level instrument utilizes ratio beam optics to give exceptional stability for a host of applications including kinetics, and offers full scanning capabilities. Routine analyses are readily carried out using the simple-to-use photometry, absorbance ratio, and time scan modes. The U-5100 also features an eco-friendly design using an ultra long-life pulsed Xe lamp and easy operation with a large LCD and standard automatic 6-cell changer.
Features
- High performance affording the best baseline stability and the lowest noise in its class
- Eco-friendly system using an ultra long-life pulsed Xe lamp
- Easy operation with a large LCD and a built-in automatic 6-cell changer
- Super compact, light-weight design that easily fits on a table top
- Standard validation function for checking performance
- Modes of operation: Scanning, Time (Kinetics), Photometry and Ratio (DNA/RNA Analysis)
| U-5100 Ratio-Beam Routine Spectrophotometer Specifications |
| Wavelength range |
190 to 1,100 nm |
| Band pass |
5 nm |
| Scan speed |
40, 100, 200, 400, 800, 1200, 2400 nm/min |
| Wavelength accuracy |
±1 nm (484.6.nm) |
Wavelength setting repeatability |
±0.5 nm |
| Baseline stability |
0.0007 Abs/h (260 nm, 2 hrs after power-on)
|
| Noise level |
0.0002 Abs or less (RMS, 260 nm, 0 Abs) |
| Baseline flatness |
±0.01 Abs (200 to 950 nm) |
| Light source |
Xenon (Xe) flash lamp |
| Detector |
Silicon photodiode x2 |
| Photometric range |
Abs: -3.000 to 3.000 Abs 0 to 300%T Conc: 0.000 to 9.999
|
| Photometric accuracy |
±0.003 Abs (0 to 0.5 Abs) ±0.005 Abs (0.5 to 1.0 Abs) |
| Dimensions |
355 mm (W) x 425 mm (D) x 235 mm (H) 14 in (W) x 17 in (D) x 9.3 in (H) |
| Weight |
Approx. 13 Kg (28.7 lbs) |
| Power requirements |
100, 115, 220, 230 or 240 V, 50/60 Hz, 60 VA |