U-HPLC Systems | LaChromUltra

A complete U-HPLC system that delivers ultra high-speed performance without compromising flexibility, reliability, or ease of use.

Hitachi's Ultra High-Speed Liquid Chromatograph, LaChromUltra,delivers ultra-fast analysis, higher resolution, increased sensitivity and the flexibility to run both conventional and ultra-high resolution LC analyses on a single system. LaChromUltra is the practical next generation LC solution that meets your laboratory's changing needs AND fits your budget.

Features:

  • 60 MPa (8700 psi) pressure max and 2.0 µm particle size columns result in ultra high-speed analysis (90% faster than conventional HPLC) and dramatically improved resolution.
  • Reduced system delay volume (as low as 266 µL) minimizes peak dispersion providing improved speed, resolution and sensitivity.
  • Reduced carryover (0.005%) improves precision, sensitivity and throughput.
  • Unique hand-tight column fitting system allows no-tools-required column changing making it very easy to change columns/methods.
  • Advanced Electronic Pulsation Isolation and Compensation (A-EPIC) provides consistent solvent flow reducing pulsation problems and improving gradient accuracy.
  • Improved detector response (10ms) ensures data integrity for ultra-narrow peak widths common in fast LC.
  • Advanced Gradient Function eliminates delay volume effect from gradient analysis improving system gradient performance and reducing analysis time.
  • Full Sequence Download Function streamlines Software/Auto Sampler interface resulting in faster cycle times and higher throughput.
  • Conventional and Ultra High-Speed LC analysis on a single system allows easy method transfer and the ultimate in flexibility.
  • Method Transfer Software Tool makes it easy to convert conventional HPLC methods to work on LCU.

For detailed information, please see the LaChromUltra brochure.

Contact us to learn more about this product. For products and services outside North America, please click here.

Modules and Specifications

The LaChromUltra system consists of the following modules:

LCU System Only

  • L-2160U Pump (2 units)
  • L-2200U Autosampler
  • L-2300 or L-2350 Column Oven
  • L-2400U UV, L-2420U UV-Vis or L-2485U FL Detector
  • Organizer/Power Supply
  • Software: Empower™ 2
  • OS: Windows XP
  • Connection: USB

Specifications:

L-2160U Pump Specifications
Liquid feed system Dual-plunger reciprocating pump (series connection, pulsation suppression)
Flow rate range 0.001 to 5.000mL/min
Max. discharge pressure 60MPa (0.001 to 2.500mL/min)
Flow rate accuracy ±2% (0.101 to 2.500mL/min)
Flow rate precision RSD 0.07% or SD 0.02min
Dimensions/weight 340 (W) × 400 (D) × 150 (H)mm, 15kg
Power supply DC24V (supplied from organizer)
Power consumption 50 VA
Gradient System
Number of solutions mixable 2 solutions
Mixing system 2-pump flow rate control
L-2200U Autosampler Specifications
Number of samples 1.5mL × 200 pcs (1mL vial and micro plate usable with optional rack)
Sample injection system Direct injection
Sample injection volume 0.1 to 50µL
Injection volume reproducibility 0.3% RSD or less (when 10µL is injected)
Carryover 0.005% or less (under specified conditions)
Withstand pressure 60MPa
Sample cooling Thermoelectric cooling rack optionally available
Dimensions/weight 340 (W) × 400 (D) × 300 (H)mm, approx. 19kg
Power supply DC24V (supplied from organizer)
Power consumption 50 VA
L-2300 Column Oven Specifications
Range [Room temp - 15 °C] to[Room temp 50°C]
Setting range 1 to 65 °C
Accuracy ± 1 °C
Safety Built-in vapor/leak sensor
Stability ± 0.1 °C
Column capacity Up to three 25-cm columns
GLP functions Serial number registration in unit memory
Dimensions/weight

340 mm (W) x 400 mm (D) x 150mm (D)

approx. 10 kg

Power supply 100 V (50/60 Hz) (organizer/AC adapter not required)
Power consumption 200 VA
L-2350 Column Oven Specifications
Range [Room temp- 15 °C] to [Room temp 65 °C]
Setting range 1 to 85 °C
Accuracy ± 1 °C
Safety Built-in vapor/leak sensor
Stability ± 0.1 °C
Column capacity Up to four 50-cm columns (10 mm ID max)
GLP functions Serial number registration in unit memory
Options 3-column selector valve
6-way 2-position valve
10-way 2-position valve
Manual injector (Rheodyne 7725i)
Dimensions/weight

210 mm (W) x 360 mm (D) x 615 mm (H)

approx. 20 kg

Power supply 100V (50/60Hz) (organizer and AC adapter not required)
Power consumption 300VA
L-2400U UV Detector
Optical system Double-beam ratio photometry system
Light source D2 lamp, Hg lamp (for wavelength check)
Flow cell 3.2µL (optical path length 5mm)
Measuring wavelength range 190 to 600nm
Wavelength accuracy ±1nm
Noise 0.6 × 10–5 AU or less (250nm, air in cell, response 2.0sec, cell length 5mm)
Drift 1.0 × 10–4 AU/hr or less (250nm, air in cell, response 2.0sec, constant temperature, cell length 5mm)
Response 0.01 to 2.0sec (7 steps)
Data sampling period 10msec (shortest)
Dimensions/weight 340 (W) × 400 (D) × 150 (H)mm, approx. 11kg
Power supply DC24V (supplied from organizer)
Power consumption 85VA
L-2420U UV-Vis Detector
Optical system Double-beam ratio photometry system
Light source D2 lamp, Hg lamp (for wavelength check)
Flow cell 3.2µL (optical path length 5mm)
Measuring wavelength range 190 to 900nm
Wavelength accuracy ±1nm
Noise 0.6 × 10–5 AU or less (250nm/600nm, air in cell, response 2.0sec, cell length 5mm)
Drift 1.0 × 10–4 AU/hr or less (250nm/600nm, air in cell, response 2.0sec, constant temperature, cell length 5mm)
Response 0.01 to 2.0sec (7 steps)
Data sampling period 10msec (shortest)
Dimensions/weight 340 (W) × 400 (D) × 150 (H)mm, approx. 11kg
Power supply DC24V (supplied from organizer)
Power consumption 85VA
L-2485UFluorescence Detector
Light source Xe lamp, Hg lamp (for wavelength check)
Measuring wavelength range Excitation side: 200 to 850nm Emission side: 250 to 900nm (For 731nm or longer, the photomultiplier needs to be changed.)
Wavelength accuracy ±3nm
Spectral bandwidth Excitation side: 15nm Emission side: 15, 30nm (switchable)
Sensitivity S/N ratio at Raman peak of water: 550 or better [ultra high-speed LC flow cell: 3µL (volume irradiated)] (baseline method) (Em: 30nm) S/N ratio at Raman peak of water: 900 or better (standard flow cell)(baseline method)(Em: 30nm)(option)
Material of wetted parts Quartz, fluorocarbon resin
Response 0.01 to 2.0sec (7 steps)
Data sampling period 10msec (shortest)
Dimensions/weight 340 (W) × 400 (D) × 300 (H)mm, approx. 23kg
Power supply 100 V (50/60 Hz) (organizer and AC adapter not required)
Power consumption 450VA
Organizer
Output DC24V, 300VA Power supply to 2 pumps, 1 auto sampler and 1 detector
Dimensions/weight 340 mm (W) x 400 mm (D) x 150mm (D), approx.10 kg
Power supply 100 V (50/60 Hz)
Power consumption 400 VA

LaChromUltra Technology

Why Ultra High-Speed LC?

LaChromUltra uses 2.0 μm particle size columns resulting in dramatically higher column efficiency compared to conventional HPLC systems. When using ≤ 2.0 μm particle sizes columns, efficiency does not diminish with increased linear velocity—resulting in both increased analysis speed and higher resolution.

Van Deemter Plot

Vertical axis: theoretical plate height (Chromatographic Efficiency)
Column length/theoretical plate number -- smaller number indicates higher column performance

Horizontal axis: linear velocity (Flow rate in the column)
The optimal flow rate range for 5 μm particle is narrow while the theoretical plate height for 2 μm particle remains very consistent at increased flow rates.

As a result, the analysis time is reduced by 90% compared to our conventional HPLC system.

LCU Result Comparison

 

 